8 Performing a Measurement
D52IB001EN-G
47
In order to take the sample out of TTK 600, proceed
as follows:
1. Switch off the heater of CCU 100.
2. Open the lid.
3. Disconnect the Pt100 connector.
4. Disconnect the Gas Connector.
5. Unscrew the fixing screw of the sample holder
with supplied Torx screw driver.
6. Take the sample holder carefully out of the fixing
block by gentle pulling on the sheat resistance
temperature sensor (right side) while lifting the
cable of the temperature sensor over the heat-
ing block (left side).
8.2.4.3 Battery Sample Holder
In order to remove the sample from the TTK 600,
proceed as follows:
1. Switch off the heater of the CCU 100.
2. Open the lid.
3. Disconnect the electrode connector from the
TTK 600 housing.
4. Disconnect the Pt100 connector.
5. Unscrew the fixing screw of the sample holder
with the supplied Torx screwdriver.
6. Carefully take the sample holder out of the fixing
block by gently pulling on the sheath resistance
temperature sensor (right side) while lifting the
cables over the heating block (right side).