11 Maintenance
68
D52IB001EN-G
4. Place the heating/cooling block on the delivered
mount as shown in the next picture
5. Remove the sample holder fixing ring by
unscrewing the two screws with the supplied
Torx driver.
6. Remove the sample holder fixation ring and the
old graphite foil.
7. Replace the old graphite foil with a new one by
placing it centrically over the heater.
Mount the instrument by reversing the above men-
tioned steps.
Summary of Contents for TTK 600
Page 2: ......
Page 74: ...Appendix B Sample Chamber Dimensions 74 D52IB001EN G Appendix B Sample Chamber Dimensions ...
Page 75: ...Appendix C Declaration of Conformity D52IB001EN G 75 Appendix C Declaration of Conformity ...
Page 77: ...Appendix D Spare Parts and Accessories D52IB001EN G 77 176383 COOLING MODE SWITCHING VALVE ...