10 Troubleshooting
64
D52IB001EN-G
2. Sensor Break (E01)
3. Loop Break (E02)
4. No water flow (E03)
5. Housing Temperature too high (E04)
cause
action
temperature
sensor cable
not
properly con-
nected
Check temperature sensor
cable and connectors.
Pt100
damaged
Visual check of Pt100 for
damage.
plug of Pt100
cable
inside the
chamber discon-
nected
Connect the plug of Pt100
cable.
TTK 600 or
CCU 100
mal-
function
Contact Anton Paar
GmbH.
cause
action
heater fuse
blown
• Check the heater resis-
tance.
--> contact Anton Paar
GmbH.
• Replace heater fuse
--> contact Anton Paar
GmbH.
set temperature
cannot be
reached in the
predefined time
interval
• Check the heater resis-
tance.
--> contact Anton Paar
GmbH.
• Check whether the cool-
ing equipment is
switched off correctly
above the threshold tem-
perature.
heater defective
Contact Anton Paar
GmbH.
heater cable
not
properly con-
nected
Check heater cable and
connectors.
cause
action
water hoses
dis-
connected
Connect the water hoses on
TTK 600 to the cooling
water circuit.
no
cooling water
Turn on the cooling water.
flow controller
disconnected
Connect the water flow con-
troller to the CCU 100.
insufficient
cool-
ing water supply
Make sure the water pres-
sure is between
2 – 3 bar and the flow rate
is 0.7 – 1.5 l/min.
flow controller
does not switch
Contact Anton Paar
GmbH
cooling ducts
inside
TTK 600
housing choked
Check the cooling water
quality (algae?).
Disconnect the water
hoses, try to blow out chok-
ing material with com-
pressed-air and rinse with
clean water.
cause
action
no cooling
of
housing
• Check if TTK 600 water
hoses are connected to
the cooling water circuit
• Check if cooling water is
turned on.
insufficient
cooling
of hous-
ing
• Check if the cooling
water flow before
TTK 600 meets the
requirements.
• Check if the cooling
water flow after TTK 600
meets the requirements.
• If the flow resistance of
the TTK 600 chamber is
too high, contact Anton
Paar GmbH.
cable to ther-
moswitch
defec-
tive
Contact Anton Paar
GmbH.