8 Performing a Measurement
D52IB001EN-G
55
8.4.4 Mounting the Transmission Sample
Holder
In order to mount the transmission sample holder
proceed as follows:
1. Dismount the blind cover on the right side of the
housing and replace it with the gas connector.
Install the gas line for the connection heater
(see chapter 4.12).
2. Insert the sample holder in the groove of the
sample holder fixing ring (1) while guiding the
cable of the Pt100 (2) behind the heating/cool-
ing block.
3. Use the Torx screw driver that was delivered
with the instrument to fix the sample holder to
the fixation ring by tightening the screw (3) until
the first click of the Torx screw driver occurs.
(corresponds to a force of 1.1 Nm)
4. Connect the Pt100 to the Pt100 plug connector
(4).
5. Connect the tube for the convection heating to
the corresponding convection heater connector
(5).
6. Switch on the heater of CCU 100.
7. Connect the outside connector of the gas con-
nector to the delivered gas hose of the instru-
ment.
8. Connect the other end of the gas hose to the
gas reservoir.
9. The optimal flow that is needed depends on the
choice of the foils. Further information can be
found in chapter 8.3 Guideline for the correct
choice of conditions and foil materials.
Fig. 57: Transmission Sample Holder
1 Sample holder fixing ring
2 Pt100
3 Fixing screw
4 Pt100 plug connector
5 heater connector
In order to prevent the detector from being damaged
by the primary beam at low 2
Θ
angles, a beam stop
is delivered with the transmission sample holder.
The beam stop can be mounted/demounted easily
as it is magnetically coupled to the housing of the in-
strument. It is simply mounted by placing it over the
corresponding guiding rods (3 in Fig. 7) on the right
backside of the instrument.
NOTICE
A separate flow controller has to be mounted in
this gas line in order to be able to adjust the gas
flow accordingly. This item is not delivered by
Anton Paar GmbH. It is recommended to use a
flow controller that allows to adjust the flow
between 0.5 and 3 L/min.
3
2
1
4
5