8 Performing a Measurement
D52IB001EN-G
49
2. Remove the two screws to open the sample
holder.
3. Remove all the pieces and foils from the sample
holder.
4. Place the first foil over the first clamping piece.
5. Add the distance piece to the stack. Different
distance pieces are available: 1 mm, 0.5 mm
and 0.25 mm.
6. Fill the central room in the distance piece with
your sample. Powders and pastes can be used.
Use a spatula to compress the sample within
the compartment.
7. Add the second foil to the stack.
8. Add the clamping piece.
9. Fix the stack with the two screws.