10 Troubleshooting
D52IB001EN-G
65
6. Insufficient Sample Cooling (air cooling and liq-
uid nitrogen cooling)
7. Communication errors with the control software
8. Sensor Spread Failure (E11)
9. No / Unknown instrument present (E98)
cause
action
no
compressed-
air supply
to
CCU 100
• Make sure compressed
air is turned on.
• Make sure all air hose
connections on the sup-
ply unit and CCU 100
are tight.
air
supply
pres-
sure
too low
Increase compressed-air
pressure to
6 bar.
insufficient
com-
pressed-air sup-
ply
to TTK 600
• Make sure compressed-
air is turned on
• Make sure all air hose
connections on
CCU 100 and TTK 600
are tight.
cryostat
clogged
with ice
Remove the compressed-
air cooling unit, heat the
sample holder inside
TTK 600 to 200 °C for one
hour and blow dry air or
nitrogen through the cryo-
stat.
In case of the LN2 cooling
equipment, remove the LN2
inlet connection from the
Dewar. Close it by the use
of the delivered plug and
wait until all parts reached
room temperature. After
that heat the sample holder
of TTK 600 to 200 °C for
one hour and blow dry air or
nitrogen through the cryo-
stat by removing the plug
and inserting a suitable gas
hose in the inlet connection.
NOTICE
Risk of damage.
The pressure of the drying gas (dry air or nitro-
gen)
must not exceed 0.4 bar
during the drying
of the cryostat.
Higher pressures could damage the heater of
the instrument.
cause
action
bad
RS 232 con-
nection
check RS 232 cable (null
modem required) and con-
nectors
wrong settings in
diffractometer
software
check settings for
COM
port
,
baud rate
,
string for-
mat
and
device address
conflicting set-
tings
in CCU 100
and diffractome-
ter software
contact your diffractome-
ter manufacturer or Anton
Paar GmbH
cause
action
Graphite foil
between sample
holder and heat-
ing/cooling block
damaged
Change the graphite foil as
described in chapter 11.3.2
Pt100 of the sam-
ple holder defect
a. Check this case by per-
forming a measurement
with another sample
holder (if possible)
b.
contact your diffrac-
tometer manufacturer or
Anton Paar GmbH
cause
action
Cables between
chamber and
CCU not prop-
erly connected
Check connections
between chamber and
CCU.