9 Putting TTK 600 out of Operation
62
D52IB001EN-G
9 Putting TTK 600 out of Operation
When you stop using TTK 600 for a short period
without removing it from the diffractometer, proceed
as follows:
1. Cool down/heat up the sample holder to room
temperature.
2. Turn off the heater with the HEATER button.
3. Remove the sample and put the sample holder
back into the sample chamber.
4. When using the liquid nitrogen cooling equip-
ment:
I. Turn off the low temperature equipment.
II. Leave the LN2 inlet connection in the Dewar until
the system has reached room temperature again. In
case that the liquid nitrogen of the Dewar has to be
refilled, proceed as follows:
1. Take out the inlet connection of the Dewar.
2. Close the tube by using the delivered tube plug
(1).
3. Refill the Dewar as described in the instruction
manual of the Dewar.
4. Remove the plug of the inlet connection.
5. Place the inlet connection in the Dewar again.
6. Start the next experiment.
TIP:
The above mentioned procedures prevent con-
densation of moisture inside the liquid nitrogen cool-
ing equipment.
7. When using the compressed-air cooling equip-
ment:
III. Turn off the compressed-air supply.
1. Turn off the cooling water.
2. Turn off CCU 100 completely.
In order to remove TTK 600 from the diffractometer
continue as follows:
1. Completely turn off CCU 100 (mains switch).
2. Disconnect all electrical cables from the sample
chamber.
3. Disconnect the cooling device.
4. Disconnect the couplings of the cooling water
hoses.
5. Dismount the sample chamber from the goniom-
eter.
6. Store the sample chamber in a safe place, pref-
erably on a stage mount offered by Anton Paar
GmbH.
7. When using the liquid nitrogen cooling equip-
ment, follow the instruction for storing the liquid
nitrogen container given in the container's
instruction manual.
NOTICE
Risk of damage.
Some parts of the Low Temperature Equipment
are covered with ice. Make sure that no water
drops on electronic or otherwise sensitive parts
and regularly wipe while it is warming up.
1