58 DECADE user manual, edition 9
Table XI. Potential settings for PAD of glycols, alcohols, aldehydes at a
platinum working electrode.
s 1 2 3
t
(ms)
20 300 100 100
E (mV)
+200
+1300
-100
Table XII. LC-EC conditions for PAD of carbohydrates.
detector
DECADE, Antec Leyden
column CarboPac
PA1,
4x250mm
flow rate
1.0 ml/min
mobile phase
200 mM NaOH
sample
disaccharides 100 nmol/l, 20 µl injection
temperature 30
o
C
flow cell
VT-03 flow cell with 3.0 mm gold working electrode
mounted with 50 µm spacer
REF
Ag/AgCl, filled with saturated KCl/AgCl
I-cell
ca. 435 nA
References
1. D.C. Johnson, D. Dobberpuhl, R. Roberts and P. Vandeberg,
Review.
Pulsed amperometric detection of carbohydrates, amines and sulphur
species in ion chromatography - the current state of research
, J.
Chromatogr. 640 (1993) 79-96
2. D.C. Johnson en W.R. LaCourse,
LC with pulsed ECD at gold and
platinum electrodes
, Anal. Chem., 62 (1990) 589A-597A
3. W.R. LaCourse en D.C. Johnson,
Optimization of waveforms for pulsed
amperometric detection of carbohydrates following separation by LC
,
Carbohydrate Research, 215 (1991) 159-178
4. W.R. LaCourse en D.C. Johnson,
Optimization of waveforms for pulsed
amperometric detection of carbohydrates based on pulsed voltammetry
,
Anal. Chem. 65 (1993) 50-55