CHAPTER 11 Troubleshooting guide
73
C H A P T E R 1 1
Troubleshooting guide
No detector response
Possible cause
Remedy
No power
Check line voltage setting, plug in power
cord
Power switch off
Turn this switch ‘on’ (at the rear panel)
Faulty fuse
Replace fuse
Divergent mains voltage
Check line voltage, see page 13
Cell disconnected, or switched off Check connection
Recorder/integrator disconnected Check connection
Fouled WE
Clean WE (see flow cell manual)
High cell current
Possible cause
Remedy
Contaminated buffer
Replace buffer, do not recycle the buffer
High WE potential
Optimise potential, if possible: use smaller
WE diameter
Salt bridge in REF not saturated
Refill with wetted KCl crystals
Retained peaks from previous
runs
Wait for elution of these (very) broad
peaks
Column is ‘bleeding’
Replace column
High amount of Fe
++
in buffer
Add EDTA to buffer, rinse metal parts with
15% HNO
3
Noisy baseline
Possible cause
Remedy
Salt bridge in REF not saturated
Refill with saturated KCl, add wetted KCl
crystals
Air bubble in REF or in cell
Remove air bubble, continuously degas
the mobile phase
Slow temperature fluctuations
Isolate detector cell, set oven temperature
Fouled WE
Clean WE (see flow cell manual)
Leaking REF or cell
Tighten connections with care
Decreased sensitivity (low S/N ratio)
Possible cause
Remedy
Fouled WE by dirty samples
Clean WE (see flow cell manual), if
possible: dilute samples