Assemblies Removal and Replacement
6-10 Removal Procedures
MG362x1A MM
PN: 10370-10386 Rev. B
6-27
Remove the Modules with Mounting Bracket
The mounting bracket for most of the modules and components are shown in
.
The following modules have a mounting bracket that attaches the module to the bottom plate.
•
Switched Filter
•
Phase Noise Amp Module (Option 23)
Tools Required
•
# 1 Phillips screwdriver
•
5/16 open end wrench
•
Anritsu 01-201 torque wrench (applies a preset 8 in. lb. torque)
Preliminary
Disconnect the power cord from the instrument. Refer to following removal procedures:
•
“Remove the Handles” on page 6-12
•
“Remove the Rear Feet” on page 6-13
•
“Remove the Covers” on page 6-14
•
“Remove the Top Sub-panel” on page 6-15
.
Procedure
.
1.
Remove all cables and connectors. Refer to
Section 6-4 “Semi-Rigid Cable Installation (Standard Model)”
Section 6-5 “SMA Cable Installation (Standard Model)” on page 6-7
2.
Remove three M3x12mm screws that attach the switched filter to the bottom tray.
3.
Remove the switched filter/bracket assembly.
4.
Remove the two M3x6 mm Screw from the switched filter.
5.
Installation is performed in reverse order.
6.
Tighten the semi-rigid connectors to 8 in. lbs.
1. Switched Filter
2. M3x6 mm Screw (x2)
3. M3x12mm (x3)
4. Mounting Bracket
Figure 6-18.
Remove Mounting Bracket
1
3
2
4
Summary of Contents for Rubidium MG362 1A Series
Page 12: ...Contents 10 PN 10370 10386 Rev B MG362x1A MM ...
Page 16: ...1 9 ESD Requirements General Information 1 4 PN 10370 10386 Rev B MG362x1A MM ...
Page 30: ...2 6 RF Deck Assemblies Functional Description 2 14 PN 10370 10386 Rev B MG362x1A MM ...
Page 66: ...3 11 FM and Phase Modulation Calibration Calibration 3 36 PN 10370 10386 Rev B MG362x1A MM ...
Page 292: ...B 5 Self Test Messages Instrument Messages B 8 PN 10370 10386 Rev B MG362x1A MM ...
Page 293: ......