Assemblies Removal and Replacement
6-10 Removal Procedures
MG362x1A MM
PN: 10370-10386 Rev. B
6-15
Remove the Top Sub-panel
Use this section when remove the top front sub-panel.
Tools Required
# 1 Phillips screwdriver
Preliminary
Disconnect the power cord from the instrument. Refer to the following procedures:
•
“Remove the Handles” on page 6-12
•
“Remove the Rear Feet” on page 6-13
•
“Remove the Covers” on page 6-14
Procedure
.
1.
Remove the M3x5 screws (x16).
1.
Lift to remove the top sub-panel.
2.
Installation is performed in reverse order..