![Anritsu Rubidium MG362 1A Series Maintenance Manual Download Page 188](http://html.mh-extra.com/html/anritsu/rubidium-mg362-1a-series/rubidium-mg362-1a-series_maintenance-manual_2945429188.webp)
6-3 Adjustment and Verification After Replacing an Assembly
Assemblies Removal and Replacement
6-4
PN: 10370-10386 Rev. B
MG362x1A MM
6-3
Adjustment and Verification After Replacing an Assembly
Refer to
to determine the required adjustments and verifications after replacing an assembly.
Table 6-2.
Calibration/Verification Rework Guide (1 of 2)
Assembly
Calibration/Verification Procedure
Section Number
B2 Processor (Data Recovered)
Preliminary Calibration
B2 Processor (Data Unrecovered) Preliminary Calibration
All Calibrations
Switched Filter Shaper Calibration
“Power Level
Accuracy and
Flatness”
RF Level Calibration and Verification
ALC Bandwidth Calibration
Frequency Modulation Calibration and Verification
AM Calibration and Verification
DRO
Reference Modules
Preliminary Calibration
Spurious Signals Verification
“Power Level
Accuracy and
Flatness”
Phase Noise Verification
“Single Sideband
Phase Noise Test”
B5 ALC
Switched Filter Shaper Calibration
“Power Level
Accuracy and
Flatness”
RF Level Calibration or Verification
ALC Bandwidth Calibration
AM Calibration and Verification
Pulse Modulation Verification
“Pulse Modulation
Tests with the
Sampling
Oscilloscope”
B4 Internal Generator (Option 27)
Frequency Modulation Calibration and Verification
AM Calibration and Verification
Pulse Modulation Verification
“Pulse Modulation
Tests with the
Sampling
Oscilloscope”
B6 YIG Module
Switched Filter or High Power
(Option 15)
Digital Down Converter
Preliminary Calibration
Switched Filter Shaper Calibration
“Power Level
Accuracy and
Flatness”
RF Level Calibration or Verification
ALC Bandwidth Calibration
Frequency Modulation Calibration and Verification
AM Calibration and Verification
Pulse Modulation Verification
Summary of Contents for Rubidium MG362 1A Series
Page 12: ...Contents 10 PN 10370 10386 Rev B MG362x1A MM ...
Page 16: ...1 9 ESD Requirements General Information 1 4 PN 10370 10386 Rev B MG362x1A MM ...
Page 30: ...2 6 RF Deck Assemblies Functional Description 2 14 PN 10370 10386 Rev B MG362x1A MM ...
Page 66: ...3 11 FM and Phase Modulation Calibration Calibration 3 36 PN 10370 10386 Rev B MG362x1A MM ...
Page 292: ...B 5 Self Test Messages Instrument Messages B 8 PN 10370 10386 Rev B MG362x1A MM ...
Page 293: ......