![Anritsu Rubidium MG362 1A Series Maintenance Manual Download Page 189](http://html.mh-extra.com/html/anritsu/rubidium-mg362-1a-series/rubidium-mg362-1a-series_maintenance-manual_2945429189.webp)
Assemblies Removal and Replacement
6-3 Adjustment and Verification After Replacing an Assembly
MG362x1A MM
PN: 10370-10386 Rev. B
6-5
Phase Noise Verification
“Single Sideband
Phase Noise Test”
Spurious Signals Verification
B4 FM/Phase Modulation Module
(Option 12)
AM Module (Option 14)
Preliminary Calibration
Frequency Modulation Calibration and Verification
Spurious Signals Verification
Phase Noise Verification
“Single Sideband
Phase Noise Test”
FEM
Switched Filter Shaper Calibration
“Power Level
Accuracy and
Flatness”
RF Level Calibration or Verification
Phase Noise Verification
“Single Sideband
Phase Noise Test”
Spurious Signals Verification
Front Panel
Regulator
Power Supply
Spurious Signals Verification
Phase Noise Verification
“Single Sideband
Phase Noise Test”
Table 6-2.
Calibration/Verification Rework Guide (2 of 2)
Assembly
Calibration/Verification Procedure
Section Number
Summary of Contents for Rubidium MG362 1A Series
Page 12: ...Contents 10 PN 10370 10386 Rev B MG362x1A MM ...
Page 16: ...1 9 ESD Requirements General Information 1 4 PN 10370 10386 Rev B MG362x1A MM ...
Page 30: ...2 6 RF Deck Assemblies Functional Description 2 14 PN 10370 10386 Rev B MG362x1A MM ...
Page 66: ...3 11 FM and Phase Modulation Calibration Calibration 3 36 PN 10370 10386 Rev B MG362x1A MM ...
Page 292: ...B 5 Self Test Messages Instrument Messages B 8 PN 10370 10386 Rev B MG362x1A MM ...
Page 293: ......