![Anritsu Rubidium MG362 1A Series Maintenance Manual Download Page 67](http://html.mh-extra.com/html/anritsu/rubidium-mg362-1a-series/rubidium-mg362-1a-series_maintenance-manual_2945429067.webp)
MG362x1A MM
PN: 10370-10386 Rev. B
4-1
Chapter 4 — Performance Verification
4-1
Introduction
This chapter contains tests that can be used to verify the performance of the series MG362x1A Synthesized
Signal Generator to specifications. These tests support all instrument models having any version of firmware
and instrument models with the following options:
•
Option 2 (mechanical step attenuator)
•
Option 3 (low phase noise and ultra hi stability)
•
Option 12 (amplitude, frequency and phase modulation, internal/external)
•
Option 13 (low phase noise and high stability)
•
Option 15 (high power output)
•
Option 26 (internal/external pulse modulation)
•
Option 27 (internal low frequency and pulse generators)
4-2
Test Records
A blank copy of a sample performance verification test record for the MG362x1A. models is provided in
. Each test record contains the model-specific variables called for by the test
procedures. It also provides a means for maintaining an accurate and complete record of instrument
performance. We recommend that you copy these pages and use them to record the results of your initial
testing of the instrument. These initial test results can later be used as benchmark values for future tests of
the same instrument.
4-3
Connector and Key Notation
The test procedures include many references to equipment interconnections and control settings. For all
MG362x1A references, specific labels are used to denote the appropriate menu key, data entry key, data entry
control, or connector (such as
RF Output
). Most references to supporting test equipment use general labels for
commonly used controls and connections (such as Span or RF Input).
4-4
Measurement Uncertainty
The test records found in
specify a measurement uncertainty. The measurement uncertainty listed
in each test record includes the best estimate of the errors contributed by the measurement, test equipment,
standards, and other correction factors (for example, calibration factors and mismatch error) based on the
suggested equipment, the equipment setup, and the prescribed test procedure. Most of the uncertainties are
type-B per the ISO/IEC TAG 4 Guide for the Expression of Uncertainty in Measurement (GUM).
Summary of Contents for Rubidium MG362 1A Series
Page 12: ...Contents 10 PN 10370 10386 Rev B MG362x1A MM ...
Page 16: ...1 9 ESD Requirements General Information 1 4 PN 10370 10386 Rev B MG362x1A MM ...
Page 30: ...2 6 RF Deck Assemblies Functional Description 2 14 PN 10370 10386 Rev B MG362x1A MM ...
Page 66: ...3 11 FM and Phase Modulation Calibration Calibration 3 36 PN 10370 10386 Rev B MG362x1A MM ...
Page 292: ...B 5 Self Test Messages Instrument Messages B 8 PN 10370 10386 Rev B MG362x1A MM ...
Page 293: ......