![Anritsu Rubidium MG362 1A Series Maintenance Manual Download Page 47](http://html.mh-extra.com/html/anritsu/rubidium-mg362-1a-series/rubidium-mg362-1a-series_maintenance-manual_2945429047.webp)
Calibration
3-9 Pulse Modulation Tests
MG362x1A MM
PN: 10370-10386 Rev. B
3-17
3-9
Pulse Modulation Tests
The pulse modulation tests verify the operation of the pulse modulation circuits in the MG362x1A. Rise time,
fall time, overshoot, and power accuracy of the pulsed RF output.
Equipment Required
The following equipment is used for pulse modulation testing.
External Pulse Reference Cal
This calibration measures the actual pulse width of the function generator output.
Note
For recording test data for this section, refer to
Table 3-14.
Pulse Modulation Test
Model Number
Critical Specification
Recommended Manufacturer/
Model
WAVEFORM GENERATOR
Up to 20MHz
KEYSIGHT, 33x Series
DCA-X OSCILLOSCOPE
N/A
KEYSIGHT, N1000A
SAMPLING HEAD
80 Giga-Sample/Seconds
KEYSIGHT, N1045B
FIXED ATTENUATOR
6 dB Attenuation
ANRITSU, SC8576
FIXED ATTENUATOR
10 dB Attenuation
ANRITSU SC8577
LOW PASS FILTER
400 MHz
RLC ELECRONICS, P/N:
F-10-400-R
Figure 3-9.
Equipment Setup for the Pulse Modulation Tests
RF Output
50 Ω
RF On/Off
Unleveled
Output: +29 dBm max
Reverse: +27 dBm, 0 VDC
FREQ
LEVEL
USER
SELECT
0
1
2
3
4
5
6
7
8
9
MG36221A
Signal Generator
20 GHz
Function Generator
Sync
Sync
Output
To PULSE
TRIG IN
From PULSE
SYNC OUT
N100A DCA-X Oscilloscope
MG362x1A (DUT)
Attenuator
Sampling Head
Summary of Contents for Rubidium MG362 1A Series
Page 12: ...Contents 10 PN 10370 10386 Rev B MG362x1A MM ...
Page 16: ...1 9 ESD Requirements General Information 1 4 PN 10370 10386 Rev B MG362x1A MM ...
Page 30: ...2 6 RF Deck Assemblies Functional Description 2 14 PN 10370 10386 Rev B MG362x1A MM ...
Page 66: ...3 11 FM and Phase Modulation Calibration Calibration 3 36 PN 10370 10386 Rev B MG362x1A MM ...
Page 292: ...B 5 Self Test Messages Instrument Messages B 8 PN 10370 10386 Rev B MG362x1A MM ...
Page 293: ......