Signal Analyzer (Continuation)
Replay function
Function outline
Captured waveforms can be replayed again by using the VSA function to read saved digitize data
Conditions for measurable
waveform data
Format: I, Q (binary format)
Combination of Span, Sampling rate, and Minimum capture sample
Span
Sampling rate
Minimum capture sample
1 kHz
2 kHz
74000 (37 s)
2.5 kHz
5 kHz
160000 (32 s)
5 kHz
10 kHz
310000 (31 s)
10 kHz
25 kHz
610000 (30.5 s)
25 kHz
50 kHz
730000 (14.6 s)
50 kHz
100 kHz
730000 (7.3 s)
100 kHz
200 kHz
730000 (3.65 s)
250 kHz
500 kHz
730000 (1.46 s)
500 kHz
1 MHz
730000 (730 ms)
1 MHz
2 MHz
730000 (365 ms)
2.5 MHz
5 MHz
730000 (146 ms)
5 MHz
10 MHz
730000 (73 ms)
10 MHz
20 MHz
730000 (36.5 ms)
18.6 MHz
20 MHz
730000 (36.5 ms)
20 MHz
25 MHz
730000 (29.2 ms)
25 MHz
50 MHz
730000 (14.6 ms)
31.25 MHz
50 MHz
730000 (14.6 ms)
50 MHz
100 MHz
730000 (7.3 ms)
62.5 MHz
100 MHz
730000 (7.3 ms)
100 MHz
200 MHz
730000 (3.65 ms)
125 MHz
200 MHz
730000 (3.65 ms)
■
MS2830A-017 Noise Figure Measurement Function
∗
Frequency
Frequency range
MS2830A-040: 30 MHz to 3.6 GHz
MS2830A-041: 30 MHz to 6 GHz
MS2830A-043: 30 MHz to 6 GHz
Frequency setting range
MS2830A-040: 10 MHz to 3.6 GHz
MS2830A-041: 10 MHz to 6 GHz
MS2830A-043: 10 MHz to 13.5 GHz
NF measurement
Measurement range
Within the frequency range
(Attenuator = 0 dB, Pre-Amp = On)
– 20 to +40 dB
Instrument uncertainty
Within the measurement range
ENR: 4 to 7 dB ±0.02 dB
ENR: 12 to 17 dB ±0.025 dB
ENR: 20 to 22 dB ±0.03 dB
GAIN measurement
Measurement range
Within the frequency range
–20 to +40 dB
Instrument uncertainty
Within the measurement range
≤0.07
Resolution bandwidth
Setting range
100 kHz to 8 MHz
Connector
Noise source
Connector: Rear Panel, BNC-J
Output Voltage: 28 ±0.5 V, Pulsed
∗
: Recommending the NC346 Series noise sources by Noisecom company
Product Brochure
l
MS2830A