Terminating 360B Source Lock Mode Opera-
tions
To terminate 360B VNA source lock mode opera-
tions, you must first return the 681XXC to local con-
trol and then turn off the SS Mode.
To return the 681XXC to local control, turn off the
360B VNA.
On the 681XXC, press
SYSTEM
, then
Reset
. This
turns off the Secure mode.
Next, press
SYSTEM
, then
Config
to access the
System Configuration Menu display.
At the System Configuration Menu,press
GPIB
.
When the Configure GPIB Menu (shown below) is
displayed, press
SS Mode
to turn the SS mode off.
Tracking
Mode
When operating in tracking mode, the 360B steers
its second local oscillator frequency and phase sig-
nal so as to phase-lock itself to the reference signal
from the 681XXC. Due to the inherent resolution of
the 360B’s frequency readout, frequency resolution
is limited to 1 kHz intervals.
Tracking Mode Configuration
In order for the 681XXC to operate with a 360B in
tracking mode, the signal generator must be operat-
ing in normal mode (SS Mode off). In addition, its
GPIB address and data terminator must match the
System Bus source address and data terminator
that are set on the 360B VNA.
To verify the GPIB address and data terminator or
to turn the SS mode off, press
SYSTEM
. At the
System Menu display, press
Config
. When the Sys-
tem Configuration Menu is displayed, press
GPIB
.
The Configure GPIB Menu (shown below) is dis-
played.
7-14
681XXC OM
USE WITH
USE WITH A 360B VECTOR
OTHER INSTRUMENTS
NETWORK ANALYZER
Summary of Contents for 680XXC
Page 3: ......
Page 11: ...Table of Contents Continued...
Page 13: ...Figure 1 1 Series 681XXC Synthesized Signal Generator...
Page 21: ......
Page 37: ......
Page 128: ......
Page 130: ......
Page 131: ...LOCAL OPERATION SAMPLE MENU MAPS MENU MAP Figure 4 1 Sample Menu Map Annotated 681XXC OM 4 5...
Page 146: ......
Page 191: ......
Page 206: ......