
681XXC OM
6-11
Signal Generator Quits During Operation
(
OPERATE
light remains on)
Trouble Description: The signal generator operates for some time,
then shuts down (
OPERATE
light remains on). After a short period, the
signal generator resumes normal operation. This is an indication that
the 681XXC has reached an excessive operating temperature.
Step 1
Check that the fan is still operating during the time that the
instrument is shut down.
q
If the fan is still operating, clean the air filter (see page
6-13).
q
If the fan is not operating, call a service technician.
LOCK ERROR
is Displayed
Trouble Description: This message is displayed in the frequency pa-
rameters area to indicate that the output frequency is not phase-
locked. It is normally caused by an internal component failure.
Step 1
Perform a self-test of the signal generator by pressing the
System Menu soft-key
Selftest
.
q
If self-test does not result in an error message(s), resume
normal operation.
q
If an error message(s) is displayed, call a service techni-
cian.
Table 6-4.
Troubleshooting (2 of 3)
OPERATOR
MAINTENANCE
TROUBLESHOOTING
Summary of Contents for 680XXC
Page 3: ......
Page 11: ...Table of Contents Continued...
Page 13: ...Figure 1 1 Series 681XXC Synthesized Signal Generator...
Page 21: ......
Page 37: ......
Page 128: ......
Page 130: ......
Page 131: ...LOCAL OPERATION SAMPLE MENU MAPS MENU MAP Figure 4 1 Sample Menu Map Annotated 681XXC OM 4 5...
Page 146: ......
Page 191: ......
Page 206: ......