6-3
TROUBLESHOOTING
Table 6-4 provides procedures for troubleshooting common malfunc-
tions encountered during operation of the signal generator. Included
are procedures for troubleshooting faults that do not produce error
messages, such as, failure to power up and unexpected shutdown.
6-10
681XXC OM
Signal Generator will not turn on
(
OPERATE
light is
OFF
)
Normal Operation: When the 681XXC is connected to the power
source, the
OPERATE
light should illuminate and the instrument
should power up.
Step 1
Disconnect the 681XXC from the power source, then check
the line fuse on the rear panel.
q
If the fuse is defective, replace (see page 6-14).
q
If the fuse is good, go to the next step.
Step 2
Check to see if power is available at the power receptacle.
q
If not, move to a working receptacle.
q
If power is available, go to the next step.
Step 3
Check the power cable.
q
If defective, replace.
q
If good, call a service technician.
Signal Generator will not turn on
(
OPERATE
light is
ON
)
Normal Operation: When the 681XXC is connected to the power
source, the
OPERATE
light should illuminate and the instrument
should power up.
q
If the
OPERATE
light illuminates but the unit fails to
power up, the 681XXC has an internal component fail-
ure. Call a service technician.
Table 6-4.
Troubleshooting (1 of 3)
OPERATOR
MAINTENANCE
TROUBLESHOOTING
Summary of Contents for 680XXC
Page 3: ......
Page 11: ...Table of Contents Continued...
Page 13: ...Figure 1 1 Series 681XXC Synthesized Signal Generator...
Page 21: ......
Page 37: ......
Page 128: ......
Page 130: ......
Page 131: ...LOCAL OPERATION SAMPLE MENU MAPS MENU MAP Figure 4 1 Sample Menu Map Annotated 681XXC OM 4 5...
Page 146: ......
Page 191: ......
Page 206: ......