
At the Configure GPIB menu, press
More
to go to
an Additional Configure GPIB Menu (below).
Press
Pwr Mtr Address
to change the address of the
power meter on the GPIB (the power meter’s default
address is 13). Enter the new address, between 1
and 30, using the cursor control key or the data en-
try key pad and the terminator key
Hz
ns
ADRS
The new GPIB address will appear on the display.
Press
Pwr Mtr Select
to select the power meter
model being used. (Supported power meters are the
Anritsu ML2437A, ML2438A, and ML4803A and
Hewlett-Packard 437B, 438A, and 70100A.)
Press
Previous Menu
to return to the main Config-
ure GPIB Menu display.
At the Configure GPIB menu, press
Previous Menu
to return to the System Configuration Menu display.
Creating a Power-Offset Table
The 681XXC must be in CW frequency mode and
fixed (non-swept) power level mode in order to cre-
ate a power-offset table for user level flatness cor-
rection.
Place the signal generator in CW frequency mode by
pressing the main menu key
CW/SWEEP
SELECT
At the resulting menu display, press
CW
. The
681XXC is now in CW frequency mode.
681XXC OM
3-69
LOCAL (FRONT
LEVELING
PANEL) OPERATION
OPERATIONS
Summary of Contents for 680XXC
Page 3: ......
Page 11: ...Table of Contents Continued...
Page 13: ...Figure 1 1 Series 681XXC Synthesized Signal Generator...
Page 21: ......
Page 37: ......
Page 128: ......
Page 130: ......
Page 131: ...LOCAL OPERATION SAMPLE MENU MAPS MENU MAP Figure 4 1 Sample Menu Map Annotated 681XXC OM 4 5...
Page 146: ......
Page 191: ......
Page 206: ......