Step 4
Connect the Master instrument
RF OUT-
PUT
and the Slave instrument
RF OUT-
PUT
to the appropriate connections on the
DUT.
Initiating
Master-Slave
Operation
The following paragraphs describe how to set up
both instruments to perform master-slave opera-
tions. Use the CW Frequency Mode menu map
(Chapter 4, Figure 4-2) to follow the menu se-
quences.
To initiate master-slave operation, turn on both
instruments and place them in CW mode. The CW
Menu (below) is displayed.
On the Master instrument, press
Master Slave
to
go to the Master-Slave Menu display (below).
This menu lets you perform the following:
q
Go to the Slave Frequencies List menu.
q
Set the dF frequency for the Slave unit.
q
Set the Slave unit’s main power level (L1).
q
Set the alternate sweep power level (L2) for
the Slave unit.
q
Turn master-slave operation on and off.
Press
Slave Freqs
to go to the Slave Frequencies
List Menu display (next page).
681XXC OM
7-5
USE WITH
MASTER-SLAVE
OTHER INSTRUMENTS
OPERATION
NOTE
Master-slave operations are always
initiated in the CW frequency
mode. Once initiated, you then can
change to a sweep frequency mode
of operation by selecting the de-
sired frequency mode on the Master
instrument.
Summary of Contents for 680XXC
Page 3: ......
Page 11: ...Table of Contents Continued...
Page 13: ...Figure 1 1 Series 681XXC Synthesized Signal Generator...
Page 21: ......
Page 37: ......
Page 128: ......
Page 130: ......
Page 131: ...LOCAL OPERATION SAMPLE MENU MAPS MENU MAP Figure 4 1 Sample Menu Map Annotated 681XXC OM 4 5...
Page 146: ......
Page 191: ......
Page 206: ......