
Erasing
Stored Setups
The front panel setups are stored in non-volatile
memory. A master reset is required to erase the con-
tents of the setups and reprogram them with default
data.
To perform a master reset, proceed as follows:
Step 1
With the 681XXC in standby, press and
hold the RF OUTPUT ON/OFF key.
Step 2
Press the LINE OPERATE/STANDBY
key to turn the instrument on.
Step 3
When the first menu is displayed (after
the start-up display), release the RF
OUTPUT ON/OFF key.
The contents of non-volatile memory have now been
erased and reprogrammed with default data.
3-90
681XXC OM
NOTE
The master reset function over-
writes all information stored in the
non-volatile memory with default
values. This includes the table of
2000 frequency/power level sets
used for the list sweep mode and
the five power-offset tables used for
the user level flatness correction
function.
LOCAL (FRONT
SAVING/RECALLING
PANEL) OPERATION
INSTRUMENT SETUPS
Summary of Contents for 680XXC
Page 3: ......
Page 11: ...Table of Contents Continued...
Page 13: ...Figure 1 1 Series 681XXC Synthesized Signal Generator...
Page 21: ......
Page 37: ......
Page 128: ......
Page 130: ......
Page 131: ...LOCAL OPERATION SAMPLE MENU MAPS MENU MAP Figure 4 1 Sample Menu Map Annotated 681XXC OM 4 5...
Page 146: ......
Page 191: ......
Page 206: ......