7-5 Starting the Program and Initializing
2300-584-R Software with the Lightning ME7808A/B/C
7-4
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
6.
After you enter your name, click
OK
to continue. Next, the program searches the GPIB bus for an
Anritsu
37000 Series VNA
instrument. If the program finds any other Anritsu VNA, the program skips those
instruments without any response required from you.
7.
If the program finds a
37000 Series VNA
, it asks you for information about the VNA that you want to use
for the verification.
8.
If there are multiple instruments on the GPIB Bus, and this instrument is not the one you want to use,
click
No
to search for another instrument on the GPIB bus.
9.
If this is the required VNA, click
Yes
.
10.
After you select the instrument to use, the
Anritsu W1 Verification Enter Test Set Serial Number
dialog box
appears. Enter the 5-digit serial number of the
37000 Series VNA
.
11.
Once the serial number is entered, click
OK
to proceed. The program queries the model and serial number
information for the
LO
and
Source
instruments. This may take a few moments. The
Anritsu 37397
Verification Setup Menu
dialog box appears.
12.
In the display below, in the
Calibration Kit Type
area, select
W Coax.
In the
Serial Number
area, enter the
5-digit serial number of the calibration kit. In the
Load Cal Kit Disk From VNA Floppy
area, select
Yes
Figure 7-4.
Searching for GPIB Bus Address 6 Anritsu VNA
Note
If your system is configured with an MS4640A/B series VNA, refer to
Software with the Lightning ME7808A/B/C”
.
Figure 7-5.
Found a 37000 VNA at GPIB Address 6
Figure 7-6.
Anritsu W1 Verification Enter Test Set Serial Number