Using the 2300-584-R Software with the ME7838A4
4-16 High-Band Calibration Port 3 – Port 4
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
PN: 10410-00286 Rev. L
4-45
3.
The
VNA Calibration: Step 2 of 5
dialog box appears.
a.
Connect the Male 2.02 mm Short #1 to VNA Port 3.
b.
Connect the Female 2.65 mm Short #2 to VNA Port 4.
c.
When ready to proceed, click
OK
.
4.
The
VNA Calibration: Step 3 of 5
dialog box appears.
a.
Connect the Male 2.65 mm Short #2 to VNA Port 3.
b.
Connect the Female 3.18 mm Short #3 to VNA Port 4.
c.
When ready to proceed, click
OK
.
Figure 4-65.
VNA Calibration: Step 2 of 5 Dialog
Figure 4-66.
VNA Calibration: Step 3 of 5 Dialog