6-4 Running the PVS Program
Using the 2300-584-R Software with the ME7828A
6-6
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
6-4
Running the PVS Program
After the VNA and PC have been configured as described above, you are ready to run the program and perform
the calibration/verification. Make sure you have the following information available:
•
Serial number of the 3656B Calibration/Verification Kit
•
Serial number of the 3739A Broadband Test Set
•
Model number of the LO Source (upper signal generator, usually an MG37022A)
•
Serial number of the LO Source (from the upper signal generator’s rear panel)
•
Model number of the RF Source (lower signal generator, usually an MG37022A)
•
Serial number of the RF Source (from the lower signal generator’s rear panel)
1.
With the equipment and software configured as described in
“Configuring the Hardware” on page 6-1
turn on the computer and allow it to boot up to Windows.
2.
Locate the
W1 Verification Software
icon (shown below) on the desktop and double-click on it.
3.
The program displays an
About
dialog box with version information. Click
OK
to continue.
4.
The
Test Operator Name
dialog box appears. Enter a user name or other identification.
5.
Click OK to continue. The program searches for a compatible VNA on the GPIB. When the VNA is found,
a confirmation dialog box appears.
Figure 6-6.
About Dialog
Figure 6-7.
Name Dialog