4-17 Application Interface – Main Menu Tab
Using the 2300-584-R Software with the ME7838A4
4-50
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
Go To Local Control Button
This command allows you access to the front panel buttons while the program is running. The instrument is in
Local Lock Out
unless this button is pushed. This prevents an inadvertent front panel button actuation from
affecting the proper operation of the program.
Restart Button
The
Restart
button command allows you to change the tested instrument, the frequency range, or the selected
verification kit. Selecting
Yes
causes the program to return to the program initialization stage.
Usually, this button is used to test for second frequency band and program execution continues in
“Restart for Alternate Frequency Band” on page 4-60
and the
Choose The Restart Method
dialog box appears.
Quit Button
Selecting
Quit
brings up a confirmation dialog box shown in
. Selecting
Yes
exits the program;
selecting
No
returns you to the previous screen.
About
Displays version information, copyright, other legal notices, and company contact information.
Run All Tests
This command button runs both the Matched Thru and Mismatched Thru tests. Program execution starts with
Section 4-19 “Matched Thru Tests” on page 4-53
. When both tests are completed, program execution returns to the
Main Menu tab
shown in
.
Repeat Last Cal
This command button re-calibrates the VNA according to the CURRENT calibration. This provides a way to
recalibrate the VNA without having to reset the program parameters in case the original calibration was
flawed for some reason. When the calibration sequence is complete, program execution continues to the
Main
Menu tab
shown previously in
Caution
Modifying the VNA settings (such as power level, bandwidth, averaging, etc.) could adversely affect
the measurement results. This should only be used to view the data or for using markers.
Figure 4-71.
Confirmation Dialog
Do you want to Quit the Program