3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
PN: 10410-00286 Rev. L
5-1
Chapter 5 — Using the 2300-584-R Software
with the ME7838E4
5-1
Introduction
This chapter describes the use of the Anritsu 2300-584-R Performance Verification Software (PVS) with the
VectorStar ME7838E4 Modular BB/mm-Wave VNA Measurement System. The ME7838E4 is based on the
VectorStar MS4647B Vector Network Analyzer equipped with Option 08x.
5-2
Required Equipment
Required VectorStar ME7838E4 Broadband System, 70 kHz to 110 GHz Instruments and
Components
The VectorStar ME7838E4 Broadband/Millimeter Wave Measurement System consists of:
•
VectorStar MS4647B VNA, 70 kHz to 70 GHz, V Connectors, equipped with Option 08x
•
MN4697C 4-port Test Set
•
3736B Master Test Set
•
3739C Broadband Test Set
•
Four (4) 3743E Millimeter-Wave (mm-Wave) Modules, 70 GHz to 110 GHz.
•
Necessary rear panel interconnect cables between the test set and the VNA.
•
Necessary front panel interconnect cables between the test set and the VNA
•
Necessary interconnect cables to the 3743E mm-Wave Modules.
5-3
Configuring the Hardware
This section describes how the various system elements are interconnected and describes the preliminary steps
required for operation of the verification software. See
“Required VectorStar ME7838E4 Broadband System,
70 kHz to 110 GHz Instruments and Components” on page 5-1
for a complete equipment list.
1.
Ensure that the VNA system is set to “Broadband”.
2.
To verify the setting, from the right side
MAIN
menu, select the broadband test set button:
•
MAIN | Application | APPLICATION | Rcvr Config | RCVR CONFIG | BB/mmWave (3738 Test Set) or
BB/mmWave (3739 Test Set)
3.
Select the
Broadband
Receiver Configuration.
4.
Connect a GPIB cable between the PC Controller and the VNA Rear Panel GPIB port labeled IEEE 488.2
GPIB.
Note
Anritsu does not support tests or verification processes for wafer probe equipment. Contact the
vendor of the wafer probe equipment if such support is desired.
Caution
Do not connect the cable to the VNA Rear Panel GPIB port labeled
Dedicated GPIB
.