7-7 Verification Control Dialog Box
2300-584-R Software with the Lightning ME7808A/B/C
7-10
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
7-7
Verification Control Dialog Box
The main menu dialog box consists of three tabbed pages for Program Control, Instrument Control, and
Verification Test Control. Each tabbed page contains buttons to control various program operations as outlined
in the following sections.
Program Control Tab
•
Restart
: This command allows you to change the instrument to test, the frequency range, or the selected
verification kit. Selecting Yes causes the program to return to the program initialization stage.
•
Quit
: Selecting Quit brings up a confirmation box. Selecting Yes exits the program; selecting No returns
you to the previous screen.
•
About
: Displays version information, copyright, other legal notices, and company contact information.
Instrument Control Tab
•
Auto Scale Display
: This command automatically scales each channel on the Anritsu Vector Network
Analyzer. It is the same as selecting a front panel channel button, and then pressing the front panel Auto
Scale button for each channel.
•
Default Display Scale
: This command sets all four channels to a scale of 15 (dB or Unit)/Div and the
reference value to 0. This allows you to see if the device is remotely or properly connected.
•
Go to Remote Control
: This command is only available after you have clicked the
Go to Local Control
button. When you push this button, the program takes control of the VNA and restores the VNA to the
same state it was in after the instrument was calibrated.
•
Go to Local Control
: This command allows you access to the front panel buttons while the program is
running. The instrument is in
Local Lock Out
unless this button is pushed. This prevents an inadvertent
front panel button actuation from affecting the proper operation of the program.
Figure 7-15.
Anritsu 37397 Verification – Main Menu Tab
Note
When the program detects the Vector Network Analyzer has finished a sweep, the program auto
scales all four channels before continuing.