Using the 2300-584-R Software with the ME7838E4
5-6 Application Interface – Setup Menu Tab
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
PN: 10410-00286 Rev. L
5-15
5.
Follow the dialog box instructions for installing the 3656B calibration coefficients into the VNA, then
click
OK
.
6.
The VNA displays an acknowledgment dialog box when the file load is complete, indicating that 10 files
were loaded. After the VNA has loaded the 10 Calibration coefficient files, select
OK
on the PC
application. The application will then prompt you to remove the USB drive from the VNA and install the
USB drive to the PC.
7.
The
Application Interface
dialog box reappears displaying the
Setup Menu
tab and
Frequency Range
area.
In the
Frequency Range
area, select either the low range (40 MHz to 65 GHz) or the high range (65 GHz
to 110 GHz). For purposes of this document, the low band is done first. The range that is not selected is
calibrated during the Restart Procedure described in
Section 5-14 “Restart for Alternate Frequency
. After selecting the frequency range, click OK. The
Verification Kit Information
area
appears.
Figure 5-24.
Install Calibration Kit Coefficients Dialog
Figure 5-25.
Install USB Drive in PC Dialog