4-14 Restart for Alternate Frequency Band
Using the 2300-584-R Software with the ME7838A4
4-40
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
12.
Enter the
Frequency Range
selection. Select the frequency range that has not been tested.
a.
If the first calibration procedure selected the
40 MHz to 65 GHz Frequency Range
, select the
65 GHz to 110 GHz Frequency Range
.
b.
If the first calibration procedure selected the
65 GHz to 110 GHz Frequency Range
, select the
40 MHz to 65 GHz Frequency Range
.
13.
Enter the
Verification Kit Type
selection.
14.
Enter the
Verification Kit Serial Number
selection.
15.
Click the
Go to USB Drive Locator
button.
16.
Click the
Verify Files on USB Drive
button.
17.
Complete the calibration and verification of the devices for the second frequency range.
18.
Save the second set of calibration data and verification results to
C:\mmdcsvc\Data
.
19.
This series of automated tests has verified that the VectorStar ME7838A4 Broadband S-parameter
measurement accuracy meets factory specifications and ends the calibration and verification procedure.
20.
The program returns to the main menu described in
Section 4-9, “Application Interface – Main Menu
.
Summary of Contents for 3656B W1