![Ankom TDF Operator'S Manual Download Page 66](http://html.mh-extra.com/html/ankom/tdf/tdf_operators-manual_2944721066.webp)
Operator’s
Manual
pg.
66
Rev
A
4/30/12
This diagnostic allows you to verify that the instrument is delivering the correct amount of fluid through the tubing. This
calibration should be done after installing new tubing. To run the Volume Calibration procedure, follow the steps below.
1.
Label
and
weigh
six
small
empty
glass
beakers.
2.
Press
the
START
button
on
the
screen
above.
The following screen is displayed.
3.
Install
extra
plastic
containers
under
all
six
IDF
Delivery
Nozzles
to
catch
any
fluid
currently
in
the
lines.
With containers in place, press the OK button. After fluid empties into the containers, the following screen is
displayed.
←
Place
extra
containers
under
IDF
Delivery
Nozzles
to
catch
pre
‐
calibration
charge.
Press
OK
when
done.
OK
Summary of Contents for TDF
Page 2: ...This page intentionally left blank ...
Page 4: ...This page intentionally left blank ...
Page 16: ...Operator s Manual pg 16 Rev A 4 30 12 This page intentionally left blank ...
Page 30: ...Operator s Manual pg 30 Rev A 4 30 12 This page intentionally left blank ...
Page 46: ...Operator s Manual pg 46 Rev A 4 30 12 This page intentionally left blank ...
Page 56: ...Operator s Manual pg 56 Rev A 4 30 12 This page intentionally left blank ...
Page 70: ...Operator s Manual pg 70 Rev A 4 30 12 This page intentionally left blank ...
Page 72: ...Operator s Manual pg 72 Rev A 4 30 12 This page intentionally left blank ...