![Ankom TDF Operator'S Manual Download Page 32](http://html.mh-extra.com/html/ankom/tdf/tdf_operators-manual_2944721032.webp)
Operator’s
Manual
pg.
32
Rev
A
4/30/12
4.
Weigh
SDF
bags.
Using a Bag Weigh Holder is critical to eliminate the effects of static electricity
during the weighing process.
Roll or fold each bag and place it in a tared Bag Weigh Holder. Place the Bag Weigh Holder in the center of a scale and
record the weight.
Because different scales have different sensitivities, the Bag Weigh Holder should
be placed in the center of the scale for best results.
5.
Weigh
Diatomaceous
Earth
(DE).
DE is used during fiber analysis to enhance the flocculation and filtration of the SDF fraction. Place ca 1 g of DE in each
of six tared tins and record the weights.
6.
Weigh
Samples.
Place 0.5
0.05 g of sample in each of six tared tins and record the weights.
7.
Install
SDF
filter
bags
on
the
ANKOM
TDF
Dietary
Fiber
Analyzer.
7.1
Remove Clamp Bars A, B, C, and D.
7.2
Pull the black Delivery Nozzle out towards you.
Clamp
Bar
C
Clamp
Bar
D
Clamp
Bar
B
Clamp
Bar
A
Summary of Contents for TDF
Page 2: ...This page intentionally left blank ...
Page 4: ...This page intentionally left blank ...
Page 16: ...Operator s Manual pg 16 Rev A 4 30 12 This page intentionally left blank ...
Page 30: ...Operator s Manual pg 30 Rev A 4 30 12 This page intentionally left blank ...
Page 46: ...Operator s Manual pg 46 Rev A 4 30 12 This page intentionally left blank ...
Page 56: ...Operator s Manual pg 56 Rev A 4 30 12 This page intentionally left blank ...
Page 70: ...Operator s Manual pg 70 Rev A 4 30 12 This page intentionally left blank ...
Page 72: ...Operator s Manual pg 72 Rev A 4 30 12 This page intentionally left blank ...