Instrument
Installation
Site
Requirements
To install and operate the
ANKOM
TDF
Dietary Fiber Analyzer you will need the following:
Adjustable wrench
Adequate power (see “Operating Environment” section of this manual)
Drain or waste container to capture waste fluids
100 psi (6.9 bar) Nitrogen Supply
Because the instrument is powered by pneumatics, do NOT run it without an
adequate Nitrogen Supply. You can run approximately 5 to 10 full assays
(IDF/SDF or TDF) with a 304 cu ft tank of nitrogen.
Instrument
Installation
Procedure
To install the ANKOM
TDF
Dietary Fiber Analyzer, follow the steps below.
1.
Unpack
and
place
the
instrument
where
it
will
be
used.
The instrument shipping container consists of three separate cardboard pieces: a top, a bottom, and a sleeve that forms the
body of the container. Remove the top and the sleeve of the instrument shipping container. Lift the instrument off of the
shipping container bottom and place it on a surface that is firm and level in an area near a Nitrogen Supply and a drain or
chemical waste container.
Waste fluids are drained from the instrument using gravity. Therefore, position the
instrument so that the drain tube can run down from the Waste Tray.
The instrument must not be subjected to excessive shock, vibration, dirt, moisture, oil, or other fluids.
Ethanol is used within this instrument. Follow both local and federal regulations for
vent hood requirements when installing this instrument.
2.
Unpack
the
attachments
and
accessories.
Within the shipping container is a cardboard box containing attachments and accessories. Open the box and verify that all
of the following items are there.
Control Panel
Bag Drying and Weighing:
2 – Desiccant Pouches, 1 – Drying Rack, 1 – Bag Weigh Holder
Summary of Contents for TDF
Page 2: ...This page intentionally left blank ...
Page 4: ...This page intentionally left blank ...
Page 16: ...Operator s Manual pg 16 Rev A 4 30 12 This page intentionally left blank ...
Page 30: ...Operator s Manual pg 30 Rev A 4 30 12 This page intentionally left blank ...
Page 46: ...Operator s Manual pg 46 Rev A 4 30 12 This page intentionally left blank ...
Page 56: ...Operator s Manual pg 56 Rev A 4 30 12 This page intentionally left blank ...
Page 70: ...Operator s Manual pg 70 Rev A 4 30 12 This page intentionally left blank ...
Page 72: ...Operator s Manual pg 72 Rev A 4 30 12 This page intentionally left blank ...