When the tube integrity check is complete, the instrument will automatically execute the IDF/SDF procedure stopping
only for faults, aborts, or manual pH measurement (if configured in step 11.10). The Status screen will show actions and
faults as they occur during the automatic operation.
12.
Manually
measure
pH.
One of the questions you are asked before starting the automated procedure is: “Check pH manually?” If you answered
“Yes” to this question, the instrument will stop during the AMG phase, display the screen below, and make a sound to
remind you to manually measure pH and adjust if necessary.
When measuring pH use a probe that can be easily rinsed to avoid loss of sample. If you add acid or base to adjust the pH,
you must mix the solution in order to get an accurate pH reading. To manually mix the solution, follow the steps below.
12.1
Move the pH probe from side to side.
12.2
Using your fingers, press the outside of the bag just above the clamp bar multiple times to mix the solution.
12.3
When you have completed the pH measurement process, press the Continue button on the status screen.
Status
Overall
Progress
Action
in
progress
IDF
Time
remaining:
AMG
Phase
XX:XX:XX
Time
remaining:
XX:XX:XX
SDF
Time
remaining:
XX:XX:XX
pH
required?
ABORT
Display
Temperatures
Yes
Check
pH
then
press
continue
xx:xx:xx
Continue
AOAC
991.43
IDF/SDF
Status
Overall
Progress
Action
in
progress
IDF
Time
remaining:
Checking
tube
integrity
Please
wait…………..
XX:XX:XX
SDF
Time
remaining:
XX:XX:XX
pH
required?
Display
Temperatures
Yes
AOAC
991.43
IDF/SDF
Summary of Contents for TDF
Page 2: ...This page intentionally left blank ...
Page 4: ...This page intentionally left blank ...
Page 16: ...Operator s Manual pg 16 Rev A 4 30 12 This page intentionally left blank ...
Page 30: ...Operator s Manual pg 30 Rev A 4 30 12 This page intentionally left blank ...
Page 46: ...Operator s Manual pg 46 Rev A 4 30 12 This page intentionally left blank ...
Page 56: ...Operator s Manual pg 56 Rev A 4 30 12 This page intentionally left blank ...
Page 70: ...Operator s Manual pg 70 Rev A 4 30 12 This page intentionally left blank ...
Page 72: ...Operator s Manual pg 72 Rev A 4 30 12 This page intentionally left blank ...