Operator’s
Manual
Rev
A
4/30/12
pg.
57
Fault
Handling
When the instrument detects a problem, it generates a fault. Some faults require the instrument to abort operations. Other faults
will allow the instrument to continue operation.
The computer clears all fault codes at the beginning of each new procedure.
However, if you start a new procedure without correcting the cause of a fault, it
will continue to re-appear.
The screen below shows the primary sections of the instrument fault screen.
The fault screen includes sections that present information that cannot be changed by the user, along with buttons that allow the
user to modify the current operations or get additional information. The back button is available for returning to the Status
Screen or the Diagnostics screen depending on the function you were doing when you entered the fault screen.
For explanation purposes, we have placed a number in each section of the screen. Your computer screen will NOT show these
numbers.
1.
Fault
description
Box #1 contains the description of a fault that has recently occurred.
2.
Reset
fault
after
it
has
been
repaired
Button #2 is only displayed when a low input pressure fault occurs. When button #2 is pressed, the fault is reset.
3.
Next
fault
Multiple faults can occur during a procedure. Pressing button #3 will walk through all of the existing faults.
List
of
Faults
#
Name
Description
Result
Correction
E1 Paddle
Temperature
Sensor Failure
One of the Paddle Sensors is reporting
erroneous information.
Process continues to run.
Failed sensor is disabled.
Replace sensor.
E2 In-line
Sensor
Failure
The device that heats the fluid before
delivery is reporting a temperature that is
lower than the specified range.
In-progress processes are
aborted.
Run diagnostics.
Replace sensor or
heater.
E3 Paddle
Heater
Over Temp
The Paddle Heater is reporting a temperature
that is higher than the specified range.
IDF process is aborted.
SDF process continues.
Run diagnostics.
Replace sensor or
heater.
←
Faults
2
3
1
Summary of Contents for TDF
Page 2: ...This page intentionally left blank ...
Page 4: ...This page intentionally left blank ...
Page 16: ...Operator s Manual pg 16 Rev A 4 30 12 This page intentionally left blank ...
Page 30: ...Operator s Manual pg 30 Rev A 4 30 12 This page intentionally left blank ...
Page 46: ...Operator s Manual pg 46 Rev A 4 30 12 This page intentionally left blank ...
Page 56: ...Operator s Manual pg 56 Rev A 4 30 12 This page intentionally left blank ...
Page 70: ...Operator s Manual pg 70 Rev A 4 30 12 This page intentionally left blank ...
Page 72: ...Operator s Manual pg 72 Rev A 4 30 12 This page intentionally left blank ...