85
160-104694 Rev 01
Version 2.2.0
The device has protection fuses on the main power supply. The fuses are seated
inside the Power Supply Group, i.e., close to the Main Switch.
Figure 9-2: Power Supply Group Hosting Main Fuses
To replace the main fuses, perform the following steps:
1. Make sure that the
Main
switch is on "O" position, i.e., switch-OFF position.
2. Disconnect the
Main Power Supply Cord
from the Generator.
3. Open the cover of the main selector using a flat-bladed screwdriver as
illustrated in Figure 9-3.
4. Pull out the “Red” fuse-holder using a flat-bladed screwdriver, as illustrated in
Figure
9
‐
2:
Power
Supply
Group
Hosting
Main
Fuses
.
5. Replace the two fuses contained in the fuse-holder with new fuses, Type T 4A L
250V – 5 x 20 mm (whenever necessary).
6. Place the fuse-holder back into the Power Supply Group and close the cover.
7. Reconnect the main power cord.
CAUTION!
This operation must be carried out by qualified technical personnel.
WARNING!
Use exclusively protection fuses of the type, current and voltage values
specified by the manufacturer and indicated on the
Device Plate Label
.
Do not proceed if the Generator does not charge or discharge Capacitors
correctly when acting on the
Charge or Discharge
button.
After the
Discharge
button is pressed, the voltage indicated by the
High-
Voltage Capacitors digital
indicator
must be lower than 70 V.
Summary of Contents for NanoKnife
Page 1: ......
Page 6: ...6 160 104694 Rev 01 Version 2 2 0...
Page 16: ...16 160 104694 Rev 01 Version 2 2 0...
Page 26: ...26 160 104694 Rev 01 Version 2 2 0...
Page 30: ...30 160 104694 Rev 01 Version 2 2 0...
Page 42: ...42 160 104694 Rev 01 Version 2 2 0 Three Probe Array Four Probe Array Five Probe Array...
Page 43: ...43 160 104694 Rev 01 Version 2 2 0 Six Probe Array Six Probe Array 10 mm Six Probe Array 15 mm...
Page 70: ...70 160 104694 Rev 01 Version 2 2 0...
Page 78: ...78 160 104694 Rev 01 Version 2 2 0...