At the end of the procedure, a pop-up panel may appear on the screen, showing events that
may have occurred during the procedure. In particular, the following error messages may
appear:
An error message indicating that the procedure between specified electrode
probes has been interrupted due to excessive current.
An error message indicating that the current measured during the procedure
between the specified electrodes was very low. In this case, verify whether the
electrode probes are correctly connected to the Generator, i.e., each electrode
probe to the proper output connector on the front panel of the Generator.
For a complete list, go to Section 8.3 Generator Error Messages.
NOTE: Overlooking the above errors may result in ineffective ablation.
To view the
Result Graphs
screen, select the "Result Graphs" tab in the upper left corner of
the screen (Figure 5-51). The Result Graphs screen consists of the
Complete Procedure
Voltage Results graphs
and
Complete Procedure Current Results graphs
for all pulses of
the procedure
.
The lower sections of the screen do not change when moving from the
Procedure Parameters screen to the Result Graphs screen.
Figure 5-51: Pulse Generation Screen with Result Graphs Sub-Screen Shown
Summary of Contents for NanoKnife
Page 1: ......
Page 6: ...6 160 104694 Rev 01 Version 2 2 0...
Page 16: ...16 160 104694 Rev 01 Version 2 2 0...
Page 26: ...26 160 104694 Rev 01 Version 2 2 0...
Page 30: ...30 160 104694 Rev 01 Version 2 2 0...
Page 42: ...42 160 104694 Rev 01 Version 2 2 0 Three Probe Array Four Probe Array Five Probe Array...
Page 43: ...43 160 104694 Rev 01 Version 2 2 0 Six Probe Array Six Probe Array 10 mm Six Probe Array 15 mm...
Page 70: ...70 160 104694 Rev 01 Version 2 2 0...
Page 78: ...78 160 104694 Rev 01 Version 2 2 0...