59
160-104694 Rev 01
Version 2.2.0
Figure 5-43: Pulse Generation Screen
The
Run Section
prepares, controls, and runs the ablation delivery. It displays several
controls and messages according to the status of the ablation progress.
Deliver Test Pulse
button launches a low-voltage test pulse to the ablation site to
confirm that the electrical pathway between the electrodes is within the operating
ranges for impedance. Once the probe test is completed, control of the ablation
is switched to the Double Foot Pedal.
ECG Synchronized
status indicator is for Pulse Timing Status listed below:
•
“ECG disabled” if 90 PPM or 240 PPM are selected.
•
“ECG synchronized” if ECG synchronization is selected and the signal is
synchronized.
•
“ECG noisy” if ECG synchronization is selected and signal is too fast.
•
“ECG no signal” if ECG synchronization is selected and signal is too
slow or not present.
The
Status Panel/Message Box
displays status messages and instructions for the
User.