54
160-104694 Rev 01
Version 2.2.0
1. Select the Edit Button below the
Procedure Spreadsheet
.
2. Select the Adjust Dist button. When the Adjust Dist
button actuated, the Probe Distance Adjuster will appear.
3. Input the desired distances between probes, see Figure 5-33, into the white boxes
of the Probe Distance Adjuster. In the example below, the spreadsheet distance
between Probe 1 and Probe 3 is 2.4 cm. If the User desires to reduce the distance
between Probe 1 and Probe 3 to 2.0 cm, click on the 1-3 text box in the Probe
Distance Adjuster and input 2.0.
Figure 5-33: Procedure Spreadsheet & Adjuster
4. After making the desired changes, select the OK Button to close
the Probe Placement Adjuster and return to the
Procedure
Spreadsheet
.
5. The probes will automatically move in the
Probe Placement Grid
to reflect the
changes inputted in
Procedure Spreadsheet
via the Probe Distance Adjuster, Figure
5-34.
Figure 5-34: Before and After Using Probe Distance Adjuster
Before Changes
After Changes