63
160-104694 Rev 01
Version 2.2.0
If the User presses "Abort" the procedure will stop (Figure 5-48). The complete and
incomplete pairs will show in the status column. The User has a choice to "Continue
Procedure" which will continue where the procedure left off, or "Stop Procedure" which will
end the procedure.
Figure 5-48: Procedure Delivery Aborted
When the User chooses to "Continue Procedure," the procedure will complete.
During the procedure, an audible indication is produced: double beep for each group of
pulses.
If ECG synchronization was selected, and the ECG signal is noisy, low, or not present
during the procedure, the system will indicate by displaying the "Pulse Timing" status.
Energy delivery will not be allowed to continue until the signal is synchronized or the
procedure can be restarted if the User returns to one of the screens that has an active
Settings
button to access the
Pulse Timing Control
screen, and selects either 90 PPM or
240 PPM. This is explained further in section 5.1.2, Information screen: Clinical Data.
When a procedure is completed, an audible indication is produced: double long beeps.
Then the following message appears in the status panel, indicating that the procedure is
completed (Figure 5-49):
"Delivery completed."
Summary of Contents for NanoKnife
Page 1: ......
Page 6: ...6 160 104694 Rev 01 Version 2 2 0...
Page 16: ...16 160 104694 Rev 01 Version 2 2 0...
Page 26: ...26 160 104694 Rev 01 Version 2 2 0...
Page 30: ...30 160 104694 Rev 01 Version 2 2 0...
Page 42: ...42 160 104694 Rev 01 Version 2 2 0 Three Probe Array Four Probe Array Five Probe Array...
Page 43: ...43 160 104694 Rev 01 Version 2 2 0 Six Probe Array Six Probe Array 10 mm Six Probe Array 15 mm...
Page 70: ...70 160 104694 Rev 01 Version 2 2 0...
Page 78: ...78 160 104694 Rev 01 Version 2 2 0...