82
160-104694 Rev 01
Version 2.2.0
Message
Possible Reasons
Actions
Unable to Deliver a
Correct Pulse
The system detected a
pulse that is too long.
Reboot the system to let the auto-test
check the system.
If the auto-test fails, call AngioDynamics
Hardware Service.
Unable to Complete
Ablation: Charge
Failure
The system detected a
charge fail during the
ablation.
Reboot the system to let the auto-test
check the system.
If the auto-test fails, call AngioDynamics
Hardware Service.
Unable to
Charge/Discharge
The system detected a
problem during the
charge or discharge of
the capacitors.
Confirm that the STOP button is not
engaged.
If "New Probe Selection" button is
active, click, then return to "Pulse
Generation screen" to reinitialize and
return to the Pulse Generation screen to
start the ablation over.
If the "Continue Procedure" and "Stop
Procedure" buttons are active, Click,
"Stop Procedure," and do not save.
Then follow the steps above.
If the system is not able to recharge or
discharge the capacitors, please call
AngioDynamics Hardware Service.
Warning! Low Current
between probes {X}-
{Y}
During the ablation a
low current has been
detected between two
probes.
Confirm probe connections, placement
and procedure parameters.
It is recommended to repeat the ablation
between the probes in question.
Warning! Ablation
Aborted between
probes {X}-{Y} due to
high current.
During the ablation,
excessive current was
detected between two
probes. The system
stopped the pulse
delivery only between
the selected probes.
When there is high current, the system
gives the User at the end of ablation, the
option to "Continue Procedure" or "Stop
Procedure" for the missing pulses. If the
User chooses "Continue Procedure,"
then the User has the following options:
1.
Reduce pulse length
2.
Reduce voltage
3.
Reduce probe exposure and treat
at 2 levels
4.
Increase probe spacing (NOTE:
when changing probe spacing,
return to the
Probe Placement
Process
screen to reset ablation
parameters per the new spacing
).
Summary of Contents for NanoKnife
Page 1: ......
Page 6: ...6 160 104694 Rev 01 Version 2 2 0...
Page 16: ...16 160 104694 Rev 01 Version 2 2 0...
Page 26: ...26 160 104694 Rev 01 Version 2 2 0...
Page 30: ...30 160 104694 Rev 01 Version 2 2 0...
Page 42: ...42 160 104694 Rev 01 Version 2 2 0 Three Probe Array Four Probe Array Five Probe Array...
Page 43: ...43 160 104694 Rev 01 Version 2 2 0 Six Probe Array Six Probe Array 10 mm Six Probe Array 15 mm...
Page 70: ...70 160 104694 Rev 01 Version 2 2 0...
Page 78: ...78 160 104694 Rev 01 Version 2 2 0...