WEIGHING CAPABILITIES
Zero point
correcting capability
(Zero)
Clears the gross to zero in compliance with a command from the ZERO keys, Control I/O, etc.
Capable of enabling or disabling operations at unstable time.
Zero correction value is backed up by a battery.
Adjustable range: Can be freely set within 1 to 30 % of weighing capacity.
Zero tracking
capability
Detects gross zero point drift and performs zero correction automatically.
Tracking time: 0.0 to 5.0 (seconds)... A range can be set freely
Tracking width: 0.0 to 4.5 (D) ... A range can be set freely
Tare subtracting
function
Clears the net weight zero in compliance with a command from the TARE key, Control I/O, etc.
Capable of enabling or disabling operations at unstable time or negative weight time.
The battery backs up a tare value.
Adjustable range: gross
≤
Weighing capacity
Stability detecting
capability
When a weighed value fluctuation amount per sampling enters a set width within a set time, a stability
mark is illuminated, judging it a stable condition.
Capable of confirming stability through Control I/O.
Stability detection time: 0.1 to 5.0 (seconds) ... A range can be set freely
Stability detection width: 0 to 9 (D) ... A range can be set freely
Digital filter function
Integrating
capability
Two digital filters connected in series
Cutoff frequency range: 0.7 to 11 (Hz)
Integrating
capability
Integrates the net weight automatically or manually.
An accumulated weight and the number of accumulated weighing times are backed up by the battery.
Accumulated weight range: 0 to 99999999
Accumulated weighing times range: 0 to 99999999
OVERALL
Zero point correction value, tare value, setpoint, accumulated weight
Memory backup by the lithium battery (Approx. 10 years)
An alarm indicator lamp is turned on when the battery is running out.
Power failure countermeasure
Data backup method
Calibration data, each function data
Writes into the non-volatile memory(EEPROM).(Up to 10,000 times)
Supply voltage
100 to 120 VAC +10%/-15%, 50 Hz or 60 Hz
200 to 240 VAC +10%/-15%, 50 Hz or 60 Hz
Power consumption
Approx. 30 VA
Working temperature and humidity range
-5 to +40 °C, 20 to 85 % R.H
Mounting method
Panel mounting type
Weight (accessories not included)
Approximately 1.3 kg
2-4 ACCESSORIES
Parts Name
Q'ty
A&D’s parts Number
Description
Fuse
1
FS-EAK-1A
EAK-1A (Time lag fuse, 1 A)
1 JI-361J016-AG
FCN361J016-AG
(FUJITSU)
Connector for Control I/O
1 JI-360C016-B
FCN360C016-B (FUJITSU)
Connector for standard serial output 1
JA-TCP0576
TCP0576
(HOSHI ELECTRIC)
Capacity plate
1
N/A
Unit label
1
N/A
Panel mount packing
1
N/A
Terminal block cover
1
N/A
Rubber leg
4
N/A