![AND Weighing Indicator AD-4401 Instruction Manual Download Page 66](http://html1.mh-extra.com/html/and/weighing-indicator-ad-4401/weighing-indicator-ad-4401_instruction-manual_2941326066.webp)
64
<EXAMPLE>
The following is a program for connecting two AD-4401s and a computer by RS-422.
Setting item
Computer
AD-4401
Baud rate
960 0 bps
RSF-03 = 5
Parity bit
Even
RSF-04 = 2
Character bit
7 bits
RSF-05 = 7
Stop bit length
1 bit
RSF-06 = 1
Terminator
CR, LF
RSF-07 = 2
Output data (Display weight)
RSF-01 = 1
Data transmitting mode ( Command)
RSF-02 = 4
RS-422/485 switching (RS-422)
RSF-08 = 1
Address number
N/A
RSF-09 = 1 (#1)
RSF-09 = 2 (#2)
10
OPEN "COM:E71NN" AS#1
RS-232C stand-by
20
PRINT #1,"@01RW"
Data request for AD-4401 #1
30
LINE INPUT #1,A1$
Data receive from AD-4401 #1
40
PRINT A1$
Displaying the data from AD-4401 #1
50
FOR I=1 TO 100: NEXT I
Waiting for the next
60
PRINT #1,"@02RW"
Data request for AD-4401 #2
70
LINE INPUT #1,A2$
Data receive from AD-4401 #2
80
PRINT A2$
Displaying the data from AD-4401 #2
90
CLOSE
100
END
9-3 OP-04 RS-232C INPUT/OUTPUTS
OP-04 provides external inputs/outputs to be connected to an external indicator, printer, personal computer, etc. It has
both RS-232C inputs/outputs and current loop outputs. Current loop outputs and RS-232C outputs are internally
connected in parallel.
INTERFACE SPECIFICATION
Signal
System
Conforms to the EIA RS-232C
Data bits
7 bits, 8 bits
Parity bits
1 bit (Even or Odd), None
Stop bits
1 bit, 2 bits
Baud rate
600, 1200, 2400, 4800, 9600, 19200 bps
Code used
ASCII
Note:
OP-04 settings are made with RSF-01 through RSF-07.
These settings are common to OP-03 (RS-422/-485).
OP-04 RS-232C I/O connector, not included. Use the one below or the compatible on the market.
D-Sub 25P Plug: HDBB-25P, Hirose
Connector cover: HDB-CTF, Hirose
Summary of Contents for Weighing Indicator AD-4401