5-3 CALIBRATION RELATED FUNCTIONS
The calibration related functions are designed to set the basic constants for the indicator. They should be carried out first
thing upon installation.
Step 1
Remove the cover from the calibration switch located at the lower left of the front panel and press the
[CAL] key (found inside). “CAL” is displayed to inform you that the indicator is entering the calibration
mode.
Step 2
Press the [ENTER] key. The indicator switches to the calibration mode and “CAL SEt” is displayed. The
right most digit starts blinking.
Step 3
Press the [] key. “CAL Fnc” is displayed.
Step 4
Press the [ENTER] key. The indicator switches to the calibration related function mode. The main
display section displays “CALF-01” and the Subdisplay section displays its set value. CALF-01 is the
function for the setting of the unit.
Step 5
Select the function number you want to set, using the [<], [>], [
∧
] , and [
∨
] keys. The display section
displays a set value for the selected function. In this example, select CALF-02 to set the decimal point
position.
Step 6
In this example; you change the decimal point position to the 10
1
digit. Press the [ENTER] key. The
Subdisplay section now starts blinking.
Step 7
Press the [] key. The Subdisplay section displays “1”.
Step 8
Press the [ENTER] key. The main display section starts blinking again and you are returned to selection
of the function number (Step 5).
Step 9
Press the [ESC] key. “CAL SET” is displayed and the settings so far are saved in the EEPROM.
Step 10
This state is the same as in Step 3. Press the [ESC] key once again. The indicator is return to the normal
mode.
The normal mode is restored and the weight value is displayed.
•
When “Err” is displayed, the value input exceeds the setting range.
•