![AND Weighing Indicator AD-4401 Instruction Manual Download Page 17](http://html1.mh-extra.com/html/and/weighing-indicator-ad-4401/weighing-indicator-ad-4401_instruction-manual_2941326017.webp)
15
Step 12
The main display section displays “CAL SPn”, the Subdisplay section displays a weight value (current
weighing capacity, set value for CALF-04), and the lowest digit of the weight value blinks. Adjust to the
calibration weight value that you have on hand, using the [<], [>], and [ V ] keys.
Step 13
Place the weight on the system, and press the [ENTER] key after the motion-detect mark is turned off.
The Subdisplay section displays “——” for about 2 seconds.
Step 14
The main display section displays “CAL End.” If “C ErrXX” has been displayed so far, there is an error
taking place. See 5-5.
Step 15
Press the [ESC] key. The main display section displays “CAL SEt” and the actual load calibration data is
saved in the EEPROM.
Step 16
This state is the same as in Step 3. Press the [ESC] key again twice. The indicator will return to the
normal mode.
The normal mode is restored and the weight value is displayed.
After calibration is finished, be sure to attach the [CAL] key cover.
4-3 SETTING A SETPOINT
The setpoint refers to a weight setting such as a target weight or final weight required for a weighing sequence. The
following describes how to set the setpoint. For the details of the thumbwheel switches, see 8-2.
The following four methods are available for setting the setpoint:
1. Setting with the key switches (Thumbwheel switch not required)
2. Setting with a 5-digit thumbwheel switch
3. Setting with a 16-digit thumbwheel switch or OP-05 setpoint unit
4. Setting with the RS-232C/-422/-485 (See 9-2, 9-3)
Select one of the above-mentioned methods in the “setpoint read mode (SPF-01)” of the general functions.
All of the setpoint settings are backed up by the memory battery and held even if the power is turned off.
The setpoint differs depending on the weighing mode. When the weighing mode (CALF-14) is changed, attach a
relevant accessory label onto the specified area of the front panel. The following describes the initial setting state and no
label is required to be attached.
Summary of Contents for Weighing Indicator AD-4401
Page 1: ...AD 4401 WEIGHING INDICATOR 1386 2A IE...
Page 42: ...40...
Page 87: ...85 APPENDIX DIMENSIONS AD 4401 OP 05...
Page 88: ...86...
Page 89: ...87...
Page 90: ...88...