8-3 STANDARD SERIAL OUTPUT (SER. OUT)
The standard serial output is a 20 mA current loop. It is located on the rear panel of the indicator, and is used to connect
a remote indicator or printer. This output does not have a power supply. Therefore, the external unit must supply the
power.
Setting of the standard serial output is done with the general functions, SIF-01 through SIF-03.
CONNECTOR FOR STANDARD SERIAL OUTPUT (INCLUDED WITH THE ACCESSORIES.)
Parts Name
Q'ty
A&D’s parts Number
Description
Connector
1 JA-TCP0576
TCP0576, Hoshi Electric
Setting
SIF- Name Default
Parameter Description
1 Displayed
weight
2 Gross
3 Net
4 Tare
5 Gross/Net/Tare
A&D standard format
6 Accumulated
weight
7 Accumulated
counts
01 Output
data
1
8
Accumulated weight/ Accumulated
counts
Accumulated data
format
1 Stream
2 Auto
02
Data
transfer
mode
1
3 Manual
1 600
bps
03 Baud
rate
2
2 2400
bps
INTERFACE SPECIFICATION
Signal System
20 mA current loop 1=20 mA, 0=0
mA
Data bit
7 bits
Parity bit
1 bit (EVEN)
Stop bit
1 bit
Code used
ASCII
Note: When connecting any unit other
than an external indicator or printer
manufactured by A & D, the device
must supply power (active current loop)
or an external power supply must be
used.