![AND Weighing Indicator AD-4401 Instruction Manual Download Page 45](http://html1.mh-extra.com/html/and/weighing-indicator-ad-4401/weighing-indicator-ad-4401_instruction-manual_2941326045.webp)
43
7-5 CHECK WEIGHING MODE
CHECK WEIGHING MODE 1 (WEIGHING MODE CALF-14 = 5)
Output Terminal
Output Condition
Zero band
Gross
≤
Zero band
Hi-Hi
Net > Hi-Hi limit
Hi
Net > Target Hi limit
Go
Target Hi limit
≥
Net
≥
Target
Weight – Lo limit
Lo
Net < Target Weight – Lo limit
Lo-Lo
Net < Lo-Lo limit
When an output condition is established, a
relevant output terminal is turned on (power
continuity with output COM).
An output terminal number can be selected
with the general functions, OUTF-01 through
OUTF-08.
The setpoint signal output is turned off
unconditionally when the operation mode is
other than the normal mode.