8-2 SETPOINT
The setpoint refers to a weight setting such as the reference value and final required for the weighing sequence.
Connect the thumbwheel switch which makes this setting to the SET POINT connector.
Connection of the setpoint differs depending on the general function, “Setpoint Read Mode (SPF-01).” It also differs
depending on the weighing mode.
THUMBWHEEL SWITCH NOT REQUIRED MODES; SPF-01 = 1
You do not have to connect anything to the SET POINT connector. (Setting upon shipment)
5-DIGIT MODE; SPF-01 = 1
When the minimum graduation is set to 10 or more, the value of the thumbwheel switch is input and multiplied by 10.
CONNECTOR FOR SETPOINT (NOT INCLUDED WITH THE ACCESSORIES.)
Parts Name
Q'ty
A&D’s parts Number
Description
Connector
1 JI-361J024-AG
FCN361J024-AU(or AG), Fujitsu
Connector cover
1 JI-360C024-B
FCN360C024-B, Fujitsu
16-DIGIT MODE; SPF-01 = 2
Depends on the weighing mode. See {CHAPTER 7. WEIGHING SEQUENCE}
When the minimum graduation is set to 10 or more, the value of the thumbwheel switch is input and multiplied by 10.
Pin #
A1 1
A2 2
A3 4
A4 8
A5 Final
10
0
A6 Final
10
1
A7 Final
10
2
A8 Final
10
3
A9 Final
10
4
B12 Shield
Only use the pins listed
above.
When the minimum
graduation is set to 10 or
more, the value of the
thumbwheel switch is input
and multiplied by 10.