![AND Weighing Indicator AD-4401 Instruction Manual Download Page 60](http://html1.mh-extra.com/html/and/weighing-indicator-ad-4401/weighing-indicator-ad-4401_instruction-manual_2941326060.webp)
58
OUTPUT DATA
There are two kinds of transmission formats; “A & D standard format” and “Accumulation data format”.
The A & D standard format has been used for our weighing indicators. It consists of two headers, data, unit, and
terminator. The Accumulation data format outputs a simple header, numerical data, unit and terminator. The following
shows their examples.
<A & D STANDARD FORMAT>
Header 1
Header2
Data (Eight digits; polarity and decimal point included)
Unit
Terminators
S T ,
G
S ,
+ 0 0 1 2 3 4 5 k g CR LF
ASCII
code
Hexadecimal
number
Description
ST [53
54]
(STable)
US [55
53]
(UnStable)
Header-1
OL
[4F 4C]
(Over Load)
GS [47
53]
(GroSs)
NT [4E
54]
(NeT)
Header-2
TR [54
52]
(TaRe)
Separator ,
[2C]
0 to 9
[30 to 39]
+ [2B]
- [2D]
SP (space)
[20]
Data (ASCII code)
• [2E]
SP SP
[20 20]
(No unit)
SP g
[20 67]
(g)
kg [6B
67]
(kg)
SP t
[20 74]
(t)
Unit
lb [6C
62]
(lb)
< Example >
Header
1
Header
2
8 Digits Including Data, Polarity, and
Decimal Point
Unit Terminator
Gross
S T ,
G S ,
+ 0 0 1 2 3 4 5 k g CR LF
Header-2 is
"GS"
Net
S T ,
N T ,
+ 0 0 1 0 0 0 0 k g CR LF
Header-2 is
"NT"
Tare
S T ,
T R ,
+ 0 0 0 2 3 4 5 k g CR LF
Header-2 is
"TR"
With
decimal
point
S T ,
G S ,
+ 0 1 2 3 . 4 5 k g CR LF
"." in numerical
data
Plus
overflow
O L , G S , + SP SP SP SP . SP SP k g CR LF
Header-1 is
"OL"
Minus
overflow
O L , G S , - SP SP SP SP . SP SP k g CR LF
Header-1 is
"OL" and
polarity is "-"
Unstable U S ,
G S ,
+ 0 1 2 3 . 4 5 k g CR LF
Header-1 is
"US"
Output
OFF data
O L , G S , + SP SP SP SP . SP SP k g CR LF
Same as plus
overflow
Summary of Contents for Weighing Indicator AD-4401
Page 1: ...AD 4401 WEIGHING INDICATOR 1386 2A IE...
Page 42: ...40...
Page 87: ...85 APPENDIX DIMENSIONS AD 4401 OP 05...
Page 88: ...86...
Page 89: ...87...
Page 90: ...88...