Accumulation data format Example
Header
-1
8 Digits Including Data, Polarity, and Decimal Point
Unit
Terminator
Accumulated
weight
T W ,
+ 0 1 2 3 4 5 6 .
7 8 k g CR LF
Header is
"TW"
Overflow
T W , + SP SP SP SP SP SP SP . SP SP k g CR LF
"SP"
value
Weight -
Overflow
T W , - SP SP SP SP SP SP SP . SP SP k g CR LF
"SP"
value,
"-" polarity
Accumulated
count
T N ,
+ 0 1 2 3 4 5 6 7
8 9 SP
SP
CR LF
Header is
"TN"
Overflow
T N , + SP SP SP SP . SP SP . SP SP SP SP CR LF
"SP"
value
*The decimal point position remains unchanged in case of overflow.
Output-per-Sampling Format Example
Setpoint
data
Weight data, Hexadecimal, 6 digits
Terminator
←⎯⎯→
←⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯→
←⎯⎯⎯⎯→
Gross/
Net
3 0 0 0 2 7 1 0 CR LF
+
Overflow
0 0 7 F F F F F CR LF
Weight data
[7FFFFF]
-
Overflow
0 1 8 0 0 0 0 0 CR LF
Weight data
[800000]
Output
Off
data
0 0 7 F F F F F CR LF
Same as
+ Overflow
0
0
1
0
1
1
0
0
Comparison result acquired from the weighing
sequence is shown in hexadecimal.
Zero
range
Under (Hi-Hi)
Over (Hi)
Full Flow (Go)
Medium Flow (Lo)
Dribble Flow (Lo-Lo)
Discharge gate open
Batch
finish
Weighing data
Weighing data is described in 6 digits (HEX) and no decimal point.
(Example)
999.9 kg ---> 00270F (HEX)
-0.1 kg ---> FFFFFF (HEX)
Note) Select 19200 bps when data is output in the Output-per-Sampling Form.